00-f5 Title and Logo 00-f6
 guide Contents
 siteinfo Site Info


Chapter 14

14-01
Some Fundamentals
14-02
Conventional Spin-Echo

14-03
Gradient Echo

14-04
Angiography

Time-of-Flight
Phase Contrast
14-05
Maximum-Intensity Projection

14-06
Saturation Reduction

14-07
Contrast-Enhanced MRA

Application
Techniques
14-08
Cardiac MR Imaging

Static Studies
Flow Studies
Clinical Applications
Advanced Techniques


14-02 Conventional Spin-Echo Pulse Sequences

In spin-echo images, fast-flowing blood appears dark, slowly moving blood ap­pears relatively bright. Slow-flowing cerebrospin fluid (CSF) appears dark on T1- weighted images and bright on T2-weighted images because (at 1.5 T) the T1 re­la­xa­tion time of blood is approximately 1200 ms, while that of CSF is 3000 ms; the T2 relaxation times are 150 ms and 500-3000 ms, respectively.

Relatively fast-moving CSF, however, behaves like blood. Figure 14-03 shows an example of fast-flowing CSF and its influence upon contrast in a T2-weighted image. Fast-flowing blood appears dark in intermediately, T1- and T2-weighted spin-echo images (Figure 14-04). This can be very helpful for the differential dia­gno­sis of aneurysms, angiomas, vascular malformations or similar diseases (Figure 14-05). Bright signal intensity where flow voids are to be expected, sup­ports the diagnosis of slow flow or thrombosis.


Figure 14-03:
Patient with hydrocephalus. On this T2- weighted SE image, cerebrospinal fluid should be bright. However, due to flow effects, the CSF in the aqueduct and the upper fourth ventricle appears dark.


Figure 14-04:
Midsagittal slice through a normal brain. SE pulse sequence, from
(a) T1-weighted through
(b) and (c) intermediately weighted to
(d) T2-weighted.
The fluid signal of CSF changes ac­cor­ding­ly, and the flowing blood in the straight si­nus stays black.


Figure 14-05:
Intermediately weighted SE image of an infant’s abdomen. The dark areas in the front are air-filled bowel loops. The dark structures in front of the spine represent rapidly moving blood in a porto-caval shunt.


Figure 14-06 explains this phenomenon for blood. In spin-echo sequences, one main effect contributes to signal intensity behavior, the time-of-flight (TOF) effect of inflowing blood. This effect originates from the movement of the blood during the time between the application of the excitation and the refocusing RF pulse. A detailed description is given elsewhere.



Figure 14-06:
Basic effects of time-of-flight flow phenomena upon signal intensities in spin-echo images. The same imaging plane is first exposed to a 90°, then to a 180° pulse. If there is no flow, a bright sig­nal will be visible. Slow flow creates a signal intensity at the bright end of the gray scale, where­as fast flow leads to low signal or no signal at all (signal void). Now all excited spins have left the imag­ing plane by the time the refocusing 180° pulse is applied. SI = signal intensity; v = velocity.


Blood can move so fast (Table 14-01), that it is not subject to both the 90° and 180° pulses of an SE experiment, but only to one of them. Whichever pulse it is, no signal will be received from the moving blood; there will be a signal void. Tur­bu­lent flow contributes to this effect.


Table 14-01:
Range of values of blood velocities in the human body.
a = artery. Derived from ⇒ Bradley 1984 + 1992, ⇒ van As.


14-03 Gradient Echo Pulse Sequences

In two-dimensional rapid imaging sequences using gradient echoes, the signal-intensity behavior is more straightforward than in SE images. Here, stationary material experiences the effect of all applied RF pulses, resulting in a signal which is only a small percentage of the equilibrium signal. In the presence of flow, the spins in the slice are replaced by spins which have not experienced any of the preceding RF pulses and therefore give a much greater signal, providing that the flow is not turbulent.

Slice thickness and profile, T1 (and thus field strength), the repetition and echo times, and other intrinsic and extrinsic factors add to the extreme com­ple­xi­ty of flow signals in MR images. Blood velocity is not the only factor influencing signal intensity. Figure 14-07 explains this.

With increasing blood velocity, signal intensity increases until it reaches a steady state (Figures 14-08).


Figure 14-07:
Two components contribute to the flow sig­nal of a gradient-echo sequence: those spins which have entered the slice before the current excitation pulse and those which have been excited by the preceding pulse and remained in the slice. The fresh spins travelling fastest carry full mag­ne­ti­za­tion and contribute most to the signal; the partially saturated spins travelling slower con­tri­bu­te less.

The signal intensity (SI) is dependent on ve­lo­ci­ty: the faster the flow, the brighter the sig­nal, as shown in Figure 14-08.



Figure 14-08:
Conceptualized flow signal intensity patterns of a GRE sequence in the steady state. v = velocity.


spaceholder 600 spaceholder 600

LogoTop
LogoBottom
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1