14-02 Conventional Spin-Echo Pulse Sequences
In spin-echo images, fast-flowing blood appears dark, slowly moving blood appears relatively bright. Slow-flowing cerebrospin fluid (CSF) appears dark on T1- weighted images and bright on T2-weighted images because (at 1.5 T) the T1 relaxation time of blood is approximately 1200 ms, while that of CSF is 3000 ms; the T2 relaxation times are 150 ms and 500-3000 ms, respectively.
Relatively fast-moving CSF, however, behaves like blood. Figure 14-03 shows an example of fast-flowing CSF and its influence upon contrast in a T2-weighted image. Fast-flowing blood appears dark in intermediately, T1- and T2-weighted spin-echo images (Figure 14-04). This can be very helpful for the differential diagnosis of aneurysms, angiomas, vascular malformations or similar diseases (Figure 14-05). Bright signal intensity where flow voids are to be expected, supports the diagnosis of slow flow or thrombosis.
![]() |
![]() |
Figure 14-03: |
![]() |
![]() |
Figure 14-04: |
![]() |
![]() |
Figure 14-05: |
Figure 14-06 explains this phenomenon for blood. In spin-echo sequences, one main effect contributes to signal intensity behavior, the time-of-flight (TOF) effect of inflowing blood. This effect originates from the movement of the blood during the time between the application of the excitation and the refocusing RF pulse. A detailed description is given elsewhere.
![]() |
Figure 14-06:
Basic effects of time-of-flight flow phenomena upon signal intensities in spin-echo images. The same imaging plane is first exposed to a 90°, then to a 180° pulse. If there is no flow, a bright signal will be visible. Slow flow creates a signal intensity at the bright end of the gray scale, whereas fast flow leads to low signal or no signal at all (signal void). Now all excited spins have left the imaging plane by the time the refocusing 180° pulse is applied. SI = signal intensity; v = velocity.
Blood can move so fast (Table 14-01), that it is not subject to both the 90° and 180° pulses of an SE experiment, but only to one of them. Whichever pulse it is, no signal will be received from the moving blood; there will be a signal void. Turbulent flow contributes to this effect.
![]() |
![]() |
Table 14-01: |
14-03 Gradient Echo Pulse Sequences
In two-dimensional rapid imaging sequences using gradient echoes, the signal-intensity behavior is more straightforward than in SE images. Here, stationary material experiences the effect of all applied RF pulses, resulting in a signal which is only a small percentage of the equilibrium signal. In the presence of flow, the spins in the slice are replaced by spins which have not experienced any of the preceding RF pulses and therefore give a much greater signal, providing that the flow is not turbulent.
Slice thickness and profile, T1 (and thus field strength), the repetition and echo times, and other intrinsic and extrinsic factors add to the extreme complexity of flow signals in MR images. Blood velocity is not the only factor influencing signal intensity. Figure 14-07 explains this.
With increasing blood velocity, signal intensity increases until it reaches a steady state (Figures 14-08).
![]() |
![]() |
Figure 14-07: |
![]() |
Figure 14-08:
Conceptualized flow signal intensity patterns of a GRE sequence in the steady state. v = velocity.