00-f5 Title and Logo 00-f6
 guide Contents
 siteinfo Site Info


Chapter 4

04-01
The Spin-Lattice (T1) Relaxation Time

04-02
T1 on the
Microscopic Scale

04-03
T1 on the
Macroscopic Scale

Partial Saturation
Inversion Recovery
04-04
The Spin-Spin (T2) Relaxation Time

04-05
T2 on the
Macroscopic Scale

Spin Echo
04-06
Practical Measurements of T1 and T2

In vitro Determination
In vivo Determination
T1 (T2) Image and
  Weighted Images

Measurements in
  Medical Diagnosis

"Biomarkers"


04-06 Practical Measurements of T1 and T2

Relaxation times can be measured in different ways with various degrees of ac­cu­ra­cy.


04-06-01 In vitro Determination

High-resolution magnetic resonance spectroscopists have measured T1 values since the middle of the last century. The in vitro measurement is done on a small samp­le, approximately 0.1-1.0 ml or slightly larger in volume, in an extremely ho­mo­ge­ne­ous magnetic field.

A variety of methods has been developed to obtain maximal precision with mi­ni­mal time consumption. Typically, 15 to 30 magnetization measurements are per­for­med on the sample for different time delays, TI in inversion-recovery ex­pe­ri­ments or TR in partial saturation experiments. Based on these results, an observed T1 value is calculated, and the error limits are usually better than 5%.

T2 can be calculated with a single multiecho sequence. The more echoes one uses, the more accurate the measurement will be.

Calculations based on fast pulse sequences (i.e., not "clean" sequences other than IR or SE) lead to rough estimates of T1, T2, T2* (and proton density) values. They might be "reproducible" when repeated, but the use of relaxation time values acquired with such pulse sequences is not advisable for scientific or clinical com­pa­ri­sons.


04-06-02 In vivo Determination

Magnet systems with larger bores allowed the examination of whole organisms, ani­mals, and people, and a more physiological determination of relaxation time values than those of excised organs or tissues. Relaxation time measurements were considered very important during the first years of commercial MR imaging. All machines were programmed to create true T1 and T2 images, based on SE and IR sequences. However, soon it became clear that relaxation time values were not the claimed invaluable addition to diagnostics.

Localization. One of the major problems of in vivo relaxation time measurements is the localization of the volume to be observed. Details of such localization tech­ni­ques are given in Chapter 6. Actual accuracy of in vivo measurements depends on the number of points acquired and the quality of localization. Localization is re­la­ti­ve­ly uncomplicated in little or non-moving organs such as the brain, but de­mand­ing and partly impossible (in particular at high fields) in organs with complex mo­ve­ment patterns such as the heart.

Relaxation time values and proton density calculation. The current most de­pend­able method used to obtain a T1 image ("T1 map"), i.e., an image whose picture elements represent pure T1 values, relies on a mathematical manipulation of se­pa­ra­te­ly obtained images with different T1 influence. Measurements are easier and more accurate at low and medium fields, because T1 values are shorter, ECG trig­ger­ing is less complicated, and artifacts less pronounced at these fields.

Typically, two to four images are used and the signals mathematically processed to calculate pure T1 values. Bearing in mind that in vivo relaxation can be mul­ti­ex­po­nen­tial, it is somewhat inadequate to perform the analysis by such a limited fit to an exponential curve. T2 images are calculated from the images of a multiecho series, e.g. CPMG. In clinical settings, usually four or eight echoes are applied.

Usually, diffusion, flow, and multiexponential decays are not taken into account in the fits and noise as well as motion artifacts add to inaccuracies.

Matrix size and slice thickness as well as partial volume effects are limiting factors in relaxation-time measurements in vivo. Partial volume effects and other factors influence the measurements. Variations within the same lesion related to vascularity, necrosis, and cell behavior (macro­scopic com­part­ment­ili­zation) contribute to the overlapping of relaxation times values. All methods re­ly­ing on slices through the examined object will have as additional error source par­tial volume effects from the edges of the slices; the only method which avoids this slice problem is the true 3D volume imaging method. Standard deviation in fitting, artifacts, and variations in the selection of volume elements by the operators are all possible sources of error (Figure 04-22).

Furthermore, similar lesions may have a more than single exponential relaxation rate, e.g., brain tumors and mul­ti­ple sclerosis plaques. This is not unexpected, considering the heterogeneous nature of tumors. Reproducibility of such measurements is also limited.


Figure 04-22:
Relaxation time measurements in vivo can be performed pixel by pixel and by regions of interest of different size.
Left: Small regions of interest covering edema (green), tumor (pink), necrosis (red), etc.
Right: Large region covering the entire tumor.


The multilayered complexity of factors and features influencing and creating relaxation time and proton distribution changes is not completely understood yet [⇒ Springer]. A simplistic view offered by Koenig suggests that water molecules can wander, by thermally-induced diffusion, rather extensively throughout the intra- and extracullular regions of tissue, and that the exploration is rather thorough in a time of the order ot T1 (or even T2). Another concept is the highly structured water, restrained for a significant time in a geometry defined by various ionic and molecular constituents of the cytoplasm [⇒ Koenig 1985, 1988].

However, some features of the T1-dispersion do not fit easily into these concepts, for instance cross relaxation phenomena that lead to quadrupolar dips in the T1-dispersion plot. They are dependent on field strength and temperature (Figure 04-23).

More about the dependence of relaxation times on static field strength and its influence upon contrast can be found in Chapter 10.




Figure 04-23:

The dispersion of T1 in tissues (ms) versus field strength (log Tesla) is not as monotonic and smooth as shown e.g. in Figure 04-04 and Figure 10-16. This nuclear magnetic relaxation dispersion (NMRD) curve with higher resolution of a multiple sclerosis tissue sample reveals two dips (quadrupolar dips) at 0.0505 and 0.0660 T (2.1 and 2.8 MHz) where the otherwise steady increase of T1 is interrupted (after ⇒ Rinck 1988).


Faster data acquisition. Precise measurements require long acquisition times; the repetition time, TR, should be equal to or greater than 5×T1. At 0.15 T, the T1 of myocardium is around 380 ms, at 1.5 T it has climbed to around 1000 ms. Such measurements at low fields take approximately 5 minutes, at high or ultra-high field more than 10, perhaps 15 minutes. Thus, faster acquisition methods were sought and developed.

Fast acquisition of quantitative T1 maps can, e.g., be based on a series of snap­shot fast low-angle shot (FLASH) images after inversion of the magnetization [⇒ Deichmann]. Such techniques were for instance used for estimating the con­cen­tra­tion of paramagnetic contrast agents in an organ.

Since the acquisition of quantitative tissue data from a beating heart has to be very fast, lately much research is focused on modifications of a pulsed NMR se­quen­ce proposed by David C. Look and Donald R. Locker in 1969. MRI did not exist at that time, and Look and Locker used their time-saving one-shot method for NMR spectroscopy instead of the conventional methods to measure the T1 re­la­xa­tion time. The spectroscopic “LL” method was within 10% of the conventionally precise-calculated value [⇒ Look+Locker].

In the 1980s, the method was further developed for MRI by Graumann and his colleagues [⇒ Graumann]. Others followed and precision was waived for speed. Among the modified sequences for cardiac and, e.g., brain MRI experimentally (and some­times cli­ni­cal­ly) used today, one finds PURR [⇒ Lee], MOLLI [⇒ Mess­rogh­li], and ShMOLLI [⇒ Piechnik]. They all suffer to varyiing extent from errors, resulting in an underestimation of the true T1. "Apparent" T1 values of MOLLI and ShMOLLI measurements of, e.g., normal myocardium have an error range of 30% or higher. A number of different pulse sequences, e.g., SASHA, SAPPHIRE, DESPOT and many others were also being tested.

Imperfect spoiling of transverse magnetization at higher flip angles in gradient echo sequences has a negative effect on a precise estimation of signal intensity and other parameters, such as relaxation times [⇒ Zur].

Critical remarks. Cardiologists are using "apparent" T1 for "cardiac mapping" and T2* (in reality T2app) for measuring cardiac iron overload, which according to them is, so far, one of the most useful ap­proa­ches of modern cardiac medicine. The sharp fall of T2* values at high and ultra-high fields is related to the drastic rise of magnetic sus­cep­ti­bi­li­ty effects which grow linearly with magnetic field strength; pure T2 is not affected in the same way.

In some publications, additional terms are also used wrongly or confusingly, for instance T1* (a term which is not appropriate because T1 is not affected by sus­cep­ti­bi­li­ty effects) for an apparent T1 (T1app or T1influx).

From a scientific point of view, these measurements are wrong because the margin of error is huge. However, in medicine to be imprecise does not preclude specific use.

spaceholder red

inkpot
Enter: the cardiologists. Welcome to the scientific arena.
A comment.

inkpot
And a follow-up: The return of MR fingerprinting.
A comment.

spaceholder red

Quantification of MR Parameters and Synthetic Images. Relaxation time and proton density values can be used to create synthetic or simulated images for training and teaching purposes.


04-06-03 T1 (T2) Image and Weighted Images

In clinical routine, people often talk about T1, T2 or proton-density images. The right terms should be T1-weighted, T2-weighted, and proton density (ρ)-weighted (or better, intermediately weighted) images, because these images have only a cer­tain T1, T2, or proton-density dependence. However, they are not calculated pure relaxation time or proton density images. Chapter 10 will explain this in detail.

Figures 04-22 c and d are T1-weighted and T2-weighted images, to be compared with the pure T1 and T2 images of Figure 04-24a and 04-24b.




Figure 04-24:

Top row: Images of a patient with a polyp in the left paranasal sinus; bottom row: images of a cervical spine.

(a) Calculated (pure) T1, and (b) calculated (pure) T2 image. Figures (c) and (d) show T1- and T2-weighted images. Pure T1 and T2 ima­ges are of very limited diagnostic value. Multiparameter weighted images are far more valuable for clinical diagnosis and commonly used in patient studies.

Simulation software: MR Image Expert®


04-06-04 Measurements in Medical Diagnosis

Fifteen years after the first description of different relaxation behavior in tissues by Erik Odeblad [⇒ Odeblad], other researchers started postulating that relaxation ti­mes differentiate tumors from normal tissue since most T1 (and in a similar way T2) values of pathologic tissue can differ markedly from the T1 of the similar nor­mal tissue [⇒ Damadian] (cf. History of MRI).

However, the ability to discriminate, type, or even grade tumors using re­la­xa­tion time values has remained a dream, despite the sophisticated multi-point fits in­tro­du­ced over the years. Figure 04-25 shows that there are differences between, in this case, T2 of normal and diseased tissues. Although values of T2 are more ac­cu­ra­te than those of T1 because more points are used for their calculation, these dif­fe­ren­ces are not significant between T2 values of, for instance, tumors and edema or infarction.


Figure 04-25:
T2 values of normal and pathological hu­man brain tissues. The standard deviation (SD) is given in light green. The SD of nor­mal tissues can reach 20%, that of pathological tissues 30% (after ⇒ Rinck 1985).


Every year, the literature announces new attempts to exploit relaxation-time measurements in vivo. There are some positive reports about its successful use. Most concern follow-up of therapy, with patients being their own reference. Pub­li­ca­tions include, for instance, the report that relaxation times from leukemic bone marrow can be used for the differential diagnosis of this disease (Figure 04-26) [⇒ Jensen]. Similar results in high-grade gliomas have been published by another research group [⇒ Boesinger].


Figure 04-26:
T1 measurements. Follow-up of treatment of acute myeloblastic leukemia. Responder: green; non-responder: red.


Another relaxation-time study is the measurement of normal appearing white brain matter in multiple sclerosis (MS) patients. Pixel-by-pixel mapping suggests that there could be minute invisible changes in the white matter, which might ex­plain brain function deficits that cannot be explained by the size and location of visible MS plaques [⇒ Barbosa; ⇒ Lacomis; ⇒ Rinck 1987].

Yet, the follow-up of treatment based upon relaxation-time values is difficult and in most instances dubious (Figure 04-27). A rise and subsequent decline of relaxation time values after a local intervention might rather indicate edema and inflammation than successful treatment [⇒ Zhang 2014].


Figure 04-27:
Relaxation-time measurements of iden­ti­cal samples under identical measurement conditions can reveal great standard de­via­tions, as shown in this example. Re­ly­ing on in vivo measurements to evaluate the outcome of treatment is dubious. Only in some instances do massive changes allow a positive assessment.


After absolute T1 and T2 values had been used unsuccessfully by researchers, com­bi­na­tions of T1 and T2, histogram techniques, and more sophisticated three di­men­sio­nal display techniques of factor representations were used [⇒ Skalej].

Availibility of databases of in vivo relaxation-time measurements is very limited. The largest collection of data was published by Bottomley et al. [⇒ Bottomley 1984, 1987]. Unfortunately, these values are not very reliable. A comparison between in vivo and in vitro relaxation measurements is quite difficult because many T1 re­la­xa­tion time values change rapidly after excision. Only brain tissues reveal a re­la­ti­ve­ly stable relaxation behavior after they have been removed from the body [⇒ Fi­scher 1989].

spaceholder 600   Critical Remarks: Biomarkers. To add confusion to complicated science, chan­ges of terms and ter­mi­no­lo­gy are common in contemporary bioscience. Many phy­sio­lo­gi­cal and bio­lo­gi­cal mea­su­re­ments and calculations are subsumed under new terms such a "ra­dio­mics", "mo­le­cu­lar imag­ing", "celluar imaging", "personal imag­ing", "pre­ci­sion imaging", and others. Thirty years after the description of re­la­xa­tion times and re­la­xa­tion ra­tes as possible or questionable biological indicators they are re-rank­ed among bio­lo­gi­cal markers or biomarkers. None of these terms describes any no­vel scientific dis­ci­pli­ne, but rather existing specialized ap­pli­ca­tions of certain mea­su­re­ment tech­ni­ques.

In general, biomarkers are biological indicators of any kind; there are thousands of them. They are not specific for MR imaging or MR spectroscopy. Typical bio­mar­kers are measurements or scores such as blood pressure, body temperature, or the body mass index.

In MR imaging, biomarkers break down into numerous subgroups where they can be applied standing alone or several combined, relaxation times being only one of them. Aside of T1 and T2, there are other possible indicators for the detection, diagnosis, and monitoring of treatment, i.e., of particular physiological or disease states. Quantification of MR parameters is discussed in Chapter 15. Biomarkers ex­tract­ed through image segmentation and multispectral analysis are also de­scrib­ed in Chapter 15, those acquired with the help of contrast agents in Chapter 13 and by dynamic imaging in Chapter 16.



spaceholder red

inkpot If these scientific details were too boring, get the
"Relaxation Times Blues":
A short excursion into the background and history of T1 and T2.

spaceholder 600 spaceholder 600

LogoTop
LogoBottom
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1