00-f5 Title and Logo 00-f6
 guide Contents
 siteinfo Site Info


Chapter 6

06-01
Composition of MR Images

06-02
Localization of Spins with Field Gradients

06-03
Excitation of Selected Spins

Spin Echo
Gradient Echo
06-04
Spatial Encoding

Frequency Encoding
Phase Encoding
06-05
Tomographic (2D) Slices

Slice Definition
Slice Selection
06-06
Multiple Slices

06-07
The Complete Imaging Experiment

Frequency Encoding
2D Fourier Transform
06-08
Partial Fourier Imaging

06-09
Three-Dimensional Fourier Imaging

06-10
Parallel Imaging


06-02 Localization of Spins with Field Gradients

In imaging experiments we are not concerned with chemical shift information, rather we require positional information – we want to know from where within our sample a certain signal originates.

In Figure 06-03 we examine three small samples of water placed at different positions along the x-axis. Without the magnetic field gradient, applying an RF pulse produces a signal consisting of a single frequency; Fourier-transforming this signal gives a spectrum with a single peak.


Figure 06-03:
The signals and spectra from three water samples at different positions along the x-axis with and without the application of a magnetic field gradient along the x-axis. In the presence of the gradient, the signals from the three samples are resolved, with the separation of the signals being dependent on their spatial separation in the x-direction and on the strength of the magnetic field gradient.
(FT = Fourier transformation).


If a magnetic field gradient is present when we measure the signal, the signal consists of three different frequencies corresponding to the three different positions. As stated before, the Larmor frequency is proportional to the magnetic field strength. If one varies the frequency of the signal by changing the magnetic field linearly across the sample, the frequencies at different locations will also vary linearly.

This technique is known as imposing a field gradient. Today, all MR imaging methods utilize such magnetic field gradients (Figure 06-04) for spatial encoding.



Figure 06-04:

Effect of field gradients: the frequency range is spread out. In this case, the gradient follows the x-direction. The frequency in the center is the ‘exact’ resonance frequency.


Fourier-transforming the signal gives a spectrum of three peaks corresponding to the three different sample positions. The frequency differences between the samples depend on their physical separation and the strength of the magnetic field gradient. At the center of the magnet, the resonance frequency is unchanged since the gradient has no effect at the center. On either side, the resonance frequency will be either higher or lower depending on the polarity of the gradient (Figure 06-05).


Figura 06-05:
Magnetic gradient fields are superimposed upon the static magnetic field. Different parts of the sample experience different magnetic field strengths. Only in the center of the sample is there no change of the static magnetic field, and thus of the resonance frequency ν0.


The (magnetic) field gradients are generated by a set of coils positioned within the magnet. They can produce fields which vary uniformly along each of the three main axes (x, y, z). These linear field gradients have a strength of up to 30 milli-Tesla per meter (mT/m) in standard clinical systems, but much stronger gradients can be obtained by using smaller gradient coils or in specialized imaging methods.

Although the frequency variations produced by the gradients are very small compared to the resonance frequency, the range of resonance frequencies created is sufficient for high-resolution MR imaging. For example, to produce a frequency distribution of 25 kHz over a distance of 30 cm requires a gradient of only 2 mT/m.

The principle of the generation of a field gradient and the shape of gradient coils in a whole-body imaging system have been explained in Figure 01-04 and Figure 03-08.

Figure 06-06 shows how a pulsed magnetic field gradient is usually depicted in pulse sequence diagrams. In this case the gradient pulse is positive. Because there are three gradient directions x, y, and z, one can find gradients depicted in three "electronic channels" in pulse sequence diagrams. Gradient pulses can consist of several different components.


Figure 06-06:
Schematic representation of a pulsed field gradient used for pulse sequence diagrams.


The amplitude of the gradient is determined by the current flowing in the gradient coils. Shortening the rise time requires a faster rate of change of the voltage in the gradient coils.


spaceholder 600 spaceholder 600

LogoTop
LogoBottom
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2