00-f5 Title and Logo 00-f6
 guide Contents
 siteinfo Site Info


Chapter 6

06-01
Composition of MR Images

06-02
Localization of Spins with Field Gradients

06-03
Excitation of Selected Spins

Spin Echo
Gradient Echo
06-04
Spatial Encoding

Frequency Encoding
Phase Encoding
06-05
Tomographic (2D) Slices

Slice Definition
Slice Selection
06-06
Multiple Slices

06-07
The Complete Imaging Experiment

Frequency Encoding
2D Fourier Transform
06-08
Partial Fourier Imaging

06-09
Three-Dimensional Fourier Imaging

06-10
Parallel Imaging


06-06 Multiple Slices

A number of possible slice configurations in multiple slice imaging is depicted in Figure 06-18. Provided there is no overlap between the slices, then each slice should be totally independent of the other slices.


If there is an overlap, the RF pulses used for excitation of the different slices interfere with each other and interrupt, for instance, the T1 re­co­ve­ry in adjacent slices, leading to reduced signal-to-noise ratio and influencing signal intensities.

Commonly, a small gap is retained between slices to avoid this kind of crosstalk, or the excitation pulses are distributed over two packages; the first one excites the odd, the se­cond one the even slices.



Figure 06-18:
Multiple slices: (a) multiple slices with wide gaps; (b) multiple contiguous slices; and (c) multiple overlapping slices.


In many pulse sequences, there is a quite long delay between each excitation (repetition time = TR) of a particular slice while the magnetization recovers. Because of the relatively long T1 relaxation time of tissues, a delay of up to three seconds may be necessary before repeating the excitation. To make the most efficient use of this time, we can excite a number of parallel slices in each in­ter­val, which is achieved by changing the frequency of the RF pulse.

This procedure can be repeated to produce a series of slices (Figure 06-19). The number of slices obtainable can be calculated by dividing the repetition time TR by the time required for each slice. For example, if TR = 400 ms and TE = 50 ms, the theoretically possible number of slices is eight (in practice seven, since each slice requires slightly more than TE). If the repetition time is long enough, not only several slices but also several images with increasing echo times per slice can be created.


Figure 06-19:

(a) Excitation of multiple slices within one TR cycle (spin-echo pulse sequence: multi-slice single echo). In this case, five slices can be excited within one repetition cycle. The excitation frequency of the individual pulses is slightly changed so that only selected nuclei, and thus slices, are excited (cf. Figure 06-17).
(b) Excitation of multiple slices within one TR cycle (SE pulse sequence). In each slice, one echo has been added to create a multi-slice double-echo sequence. The number of slices has been reduced to three because of the time restriction.


This is known as multi-slice multi-echo sequence. Commonly in an in­ves­ti­ga­tion of the brain, 15 or 16 parallel slices in the transverse view with two echoes are acquired. The repetition time (TR) is between 2,000 and 3,000 ms, the times of the two echoes (TE) are 20 and 80 ms.

Multiple slice imaging is not restricted to spin-echo sequences, but can be performed with practically all pulse sequences. One can add inversion pulses to create inversion-recovery images with a relatively long inversion time (TI); during this time, additional slices can be inverted. However, usually the number of slices is limited in IR sequences.

spaceholder 600 spaceholder 600

LogoTop
LogoBottom
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4